Riata lead failure manifests as electrical storm with lead parameters showing unique postural variations: a case report.


ABSTRACT:

Background

Riata implantable cardioverter-defibrillator (ICD) leads are prone to a unique type of mechanical lead failure causing conductor externalization (CE) which may be complicated by a delayed-onset electrical lead failure (ELF).

Case summary

A 60-year-old male with symptomatic, severe ischaemic cardiomyopathy, and atrial fibrillation following a prior anterior wall myocardial infarction received a dual-chamber ICD with 7F-RiataST ventricular lead as a primary prevention strategy against sudden cardiac death in 2008. In 2017, a pulse generator replacement was performed for elective replacement indicator status. At that time, CE was noted in the ventricular lead but the electrical lead parameters were normal, hence lead replacement was decided against and the patient wa