Preparation of Drager Atlan A350 and General Electric Healthcare Carestation 650 anesthesia workstations for malignant hyperthermia susceptible patients.


ABSTRACT:

Background

Patients at risk of malignant hyperthermia need trigger-free anesthesia. Therefore, anesthesia machines prepared for safe use in predisposed patients should be free of volatile anesthetics. The washout time depends on the composition of rubber and plastic in the anesthesia machine. Therefore, new anesthesia machines should be evaluated regarding the safe preparation for trigger-free anesthesia. This study investigates wash out procedures of volatile anesthetics for two new anesthetic workstations: Dräger Atlan A350 and General Electric Healthcare (GE) Carestation 650 and compare it with preparation using activated charcoal filters (ACF).

Methods

A Dräger Atlan and a Carestation 650 were contaminated with 4% sevoflurane for 90 min. The machines were decontaminated