A Study of Analytical and Clinical Sensitivity of Aptima SARS-CoV-2 Assay (Hologic) and Proposals of Complementary Tests for SARS-CoV-2 Detection in Low Viral Load Specimens.


ABSTRACT: Early and accurate detection of SARS-CoV-2 is important for diagnosis and transmission control. The use of high-throughput and automated testing allows laboratories to better deliver diagnostic testing given manpower and resource limitations. We validated the clinical and analytical performance of the Hologic Panther Aptima SARS-CoV-2 assay with an emphasis on detection of specimens with low viral loads. The clinical performance was evaluated using 245 clinical specimens, against a comparator PCR-based laboratory developed test (LDT). The analytical performance was determined by replicate testing of contrived samples in a ten-fold dilution series (CT values 32-42, based on LDT). The Aptima assay had 96.7% overall percent agreement, 100% negative percent agreement and 88.1% posit