Unknown

Dataset Information

Detecting Progression in Advanced Glaucoma: Are Optical Coherence Tomography Global Metrics Viable Measures?


ABSTRACT:

Significance

Optical coherence tomography (OCT) summary measures have been suggested as a way to detect progression in eyes with advanced glaucoma. Here, we show that these measures have serious flaws largely due to segmentation errors. However, inspection of the images and thickness maps can be clinically useful.

Purpose

This study aimed to test the hypothesis that recently suggested global OCT measures for detecting progression in eyes with advanced progression are seriously affected by segmentation mistakes and other errors that limit their clinical utility.

Methods

Forty-five eyes of 38 patients with a 24-2 mean deviation worse than -12 dB had at least two spectral domain OCT sessions (0.8 to 4.4 years apart) with 3.5-mm circle scans of the disc and cube scans cen

SUBMITTER: Thenappan A 

PROVIDER: S-EPMC8673437 | biostudies-literature | 2021 May