Ontology highlight
ABSTRACT: Background
We established high-sensitivity cardiac troponin I (hsTnI) 99th percentile upper reference limits (URLs) for the Centaur XPT High-Sensitivity Troponin I assay (Centaur hsTnI; Siemens, Erlangen, Germany) and Atellica IM High-Sensitivity Troponin I assay (Atellica hsTnI; Siemens) and assessed the effect of outlier elimination.Methods
The reference population comprised 380 men and 387 women, satisfying the strict systematic reference population criteria. After reference population verification by the N-terminal pro-B-type natriuretic peptide (NT-proBNP) assay, 99th percentile URLs for Centaur hsTnI and Atellica hsTnI were calculated before and after outlier elimination.Results
The 99th percentile URL for Centaur hsTnI was 60.4 (men, 74.7; women, 57.5) ng/L a
SUBMITTER: Ahn S
PROVIDER: S-EPMC8677476 | biostudies-literature | 2022 May
REPOSITORIES: biostudies-literature