Ontology highlight
ABSTRACT: Introduction
Radiofrequency catheter ablation is a common treatment for atrial fibrillation (AF), during which thermal esophageal injury may rarely occur and lead to an atrio-esophageal fistula. Therefore, we studied the utility of the Circa S-Cath™ multi-sensor luminal esophageal temperature (LET) probe to prevent esophageal thermal injury.Methods and results
Thirty-six patients, enrolled prospectively, underwent circumferential or segmental pulmonary vein isolation for treatment of AF. A maximum ablation electrode temperature of 42ºC was programmed for automatic power delivery cutoff. In addition, energy delivery was manually discontinued when the maximum LET on any sensor of the probe rose abruptly (i.e. ˃0.2ºC) or exceeded 39º C. Esophagoscopy was performed immediately
SUBMITTER: Bhuta S
PROVIDER: S-EPMC8691294 | biostudies-literature | 2020 Dec