Respiration monitoring in PACU using ventilation and gas exchange parameters.
Ontology highlight
ABSTRACT: The importance of perioperative respiration monitoring is highlighted by high incidences of postoperative respiratory complications unrelated to the original disease. The objectives of this pilot study were to (1) simultaneously acquire respiration rate (RR), tidal volume (TV), minute ventilation (MV), SpO2 and PETCO2 from patients in post-anesthesia care unit (PACU) and (2) identify a practical continuous respiration monitoring method by analyzing the acquired data in terms of their ability and reliability in assessing a patient's respiratory status. Thirteen non-intubated patients completed this observational study. A portable electrical impedance tomography (EIT) device was used to acquire RREIT, TV and MV, while PETCO2, RRCap and S
SUBMITTER: Kang HY
PROVIDER: S-EPMC8692466 | biostudies-literature | 2021 Dec
REPOSITORIES: biostudies-literature