Validation of a rapid liquid chromatography tandem mass spectrometric method for the quantitative analysis of vistusertib.
Ontology highlight
ABSTRACT: Vistusertib is an orally bioavailable mTOR inhibitor that is being studied in clinical trials. A novel reliable method was developed to quantitate vistusertib using LC-MS/MS to explore drug exposure-response relationships. Sample preparation involved protein precipitation using acetonitrile. Separation of vistusertib and the internal standard, AZD8055, was achieved with a Waters Acquity UPLC BEH C18 column utilizing isocratic elution over a 3 min total analytical run time. A SCIEX 4500 triple quadrupole mass spectrometer operated in positive electrospray ionization mode was used for the detection of vistusertib. The assay range was 5-5000 ng/mL and proved to be accurate (98.7-105.7%) and precise (CV ≤ 10.5%). A 40,000 ng/mL sample that was diluted 1:10 (v/v) with plasma was accurately quan
SUBMITTER: Scott SC
PROVIDER: S-EPMC8693366 | biostudies-literature | 2022 Jan
REPOSITORIES: biostudies-literature
ACCESS DATA