Blood Pumps for Extracorporeal Membrane Oxygenation: Platelet Activation During Different Operating Conditions.


ABSTRACT: Extracorporeal membrane oxygenation (ECMO) is a therapy used in severe cardiopulmonary failure. Blood is pumped through an artificial circuit exposing it to nonphysiologic conditions, which promote platelet activation and coagulation. Centrifugal pumps used at lower flow rates than their design point may lose pump efficiency and increase the risk of hemolysis. In this study, thrombogenic properties of two ECMO pumps designed for adult and neonatal use were evaluated using simulations in different flow scenarios. Three scenarios, adult pump in adult mode (4 L/min), adult pump in baby mode (300 ml/min), and neonatal pump used in its design point (300 ml/min), were simulated using computational fluid dynamics. The flow was numerically seeded with platelets, whose activation state was computed considering the stress history that acted along their respective path lines. Statistical distributions of activation state and residence time were drawn. The results showed that using the adult pump in baby mode increased the fraction of platelets with higher activation state confirming that low-pump flow rate impacts thrombogenicity. The neonatal pump showed a backflow at the inlet, which carried platelets in a retrograde motion contributing to an increased thrombogenic potential compared with the adult mode scenario.

SUBMITTER: Fiusco F 

PROVIDER: S-EPMC8700320 | biostudies-literature | 2022 Jan

REPOSITORIES: biostudies-literature

altmetric image

Publications

Blood Pumps for Extracorporeal Membrane Oxygenation: Platelet Activation During Different Operating Conditions.

Fiusco Francesco F   Fiusco Francesco F   Broman Lars Mikael LM   Prahl Wittberg Lisa L  

ASAIO journal (American Society for Artificial Internal Organs : 1992) 20220101 1


Extracorporeal membrane oxygenation (ECMO) is a therapy used in severe cardiopulmonary failure. Blood is pumped through an artificial circuit exposing it to nonphysiologic conditions, which promote platelet activation and coagulation. Centrifugal pumps used at lower flow rates than their design point may lose pump efficiency and increase the risk of hemolysis. In this study, thrombogenic properties of two ECMO pumps designed for adult and neonatal use were evaluated using simulations in differen  ...[more]

Similar Datasets

| S-EPMC6143512 | biostudies-literature
2024-10-15 | GSE182609 | GEO
| S-EPMC6694790 | biostudies-literature
2019-12-31 | GSE93101 | GEO
2024-10-15 | GSE182608 | GEO
2024-10-15 | GSE182600 | GEO
| S-EPMC7007856 | biostudies-literature
| S-EPMC3876455 | biostudies-literature
| S-EPMC7433883 | biostudies-literature
| S-EPMC8444511 | biostudies-literature