Perioperative Sensor and Algorithm Programming in Patients with Implanted ICDs and Pacemakers for Cardiac Resynchronization Therapy.
Ontology highlight
ABSTRACT: ICDs and pacemakers for cardiac resynchronization therapy (CRT) are complex devices with different sensors and automatic algorithms implanted in patients with advanced cardiac diseases. Data on the perioperative management and outcome of CRT carriers undergoing surgery unrelated to the device are scarce. Data from 198 CRT device carriers (100 with active rate responsive sensor) were evaluated regarding perioperative adverse (device-related) events (A(D)E) and lead parameter changes. Thirty-nine adverse observations were documented in 180 patients during preoperative interrogation, which were most often related to the left-ventricular lead and requiring intervention/reprogramming in 22 cases (12%). Anesthesia-related events occurred in 69 patients. There was no ADE for non-cardiac surgery a
SUBMITTER: Niedermeier A
PROVIDER: S-EPMC8705781 | biostudies-literature | 2021 Dec
REPOSITORIES: biostudies-literature