Cross-validation of the high-capacity tensiometer and thermocouple psychrometer for continuous monitoring of xylem water potential in saplings.


ABSTRACT: The pressure chamber, the most popular method used to measure xylem water potential, is a discontinuous and destructive technique and is therefore not suitable for automated monitoring. Continuous non-destructive monitoring could until very recently be achieved only by use of the thermocouple psychrometer (TP). Here we present the high-capacity tensiometer (HCT) as an alternative method for continuous non-destructive monitoring. This provided us with a unique chance to cross-validate the two instruments by installing them simultaneously on the same sapling stem. The HCT and the TP showed excellent agreement for xylem water potential less than -0.5 MPa. Response to day/night cycles and watering was remarkably in phase, indicating excellent response time of both instruments despite substanti