An evaluation of the diagnostic performance characteristics of the Yellow Fever IgM immunochromatographic rapid diagnostic test kit from SD Biosensor in Ghana.
Ontology highlight
ABSTRACT: Yellow fever is endemic in Ghana and outbreaks occur periodically. The prodromal signs due to Yellow Fever Virus (YFV) infection are non-specific, making clinical signs unreliable as the sole criteria for diagnosis. Accurate laboratory confirmation of suspected yellow fever cases is therefore vital in surveillance programs. Reporting of ELISA IgM testing results by laboratories can delay due to late arrival of samples from the collection sites as well as limited availability of ELISA kits. In this study, the diagnostic performance characteristics of a rapid immunochromatographic Standard Q Yellow Fever IgM test kit (SD Biosensor) was evaluated for the rapid diagnosis of Yellow Fever infection in Ghana. A panel of 275 sera, comprising 81 confirmed YFV positives and 194 negatives were re-tes
SUBMITTER: Ofosu-Appiah LH
PROVIDER: S-EPMC8741057 | biostudies-literature | 2022
REPOSITORIES: biostudies-literature
ACCESS DATA