Data on the frequency and causes of icteric interference in clinical chemistry laboratory tests.
Ontology highlight
ABSTRACT: The presence of elevated levels of bilirubin (icterus) in serum or plasma specimens has the potential to interfere with clinical chemistry and other laboratory assays. Along with hemolysis and lipemia, icterus represents one of the most common endogenous interferences with laboratory tests. There are two common mechanisms by which icterus can cause assay interference. The first common mechanism is spectral interference due to absorption at wavelengths used in assays by bilirubin and/or bilirubin breakdown products. The second common mechanism involves chemical reaction of bilirubin with the reagents used in some enzymatic assays. Most automated clinical chemistry platforms can perform rapid estimates of indices for hemolysis, icterus, and lipemia (HIL), typically by measuring absorbance at