Clinical Assessment of a Novel Ring Gantry Linear Accelerator-Mounted Helical Fan-Beam kVCT System.


ABSTRACT:

Purpose

To assess clinically relevant image quality metrics (IQMs) of helical fan beam kilovoltage (kV) fan beam computed tomography (CT).

Methods and materials

kVCT IQMs were evaluated on an Accuray Radixact unit equipped with helical fan beam kVCT to assess the capabilities of this newly available modality. kVCT IQMs were evaluated and compared to a kVCT simulator and linear accelerator-based cone beam CTs (CBCT) using a commercial CBCT image quality phantom. kVCTs were acquired on the Accuray Radixact for all combinations of kVp and mAs in fine mode using a 440-mm field of view (FOV). Evaluated IQMs were spatial resolution, overall uniformity, subject contrast, contrast-to-noise ratio (CNR), and effective slice thickness. Imaging dose was assessed for planar kV imaging.