Feasibility of MR fingerprinting using a high-performance 0.55 T MRI system.


ABSTRACT:

Background

MR fingerprinting (MRF) is a versatile method for rapid multi-parametric quantification. The application of MRF for lower MRI field could enable multi-contrast imaging and improve exam efficiency on these systems. The purpose of this work is to demonstrate the feasibility of 3D whole-brain T1 and T2 mapping using MR fingerprinting on a contemporary 0.55 T MRI system.

Materials and methods

A 3D whole brain stack-of-spirals FISP MRF sequence was implemented for 0.55 T. Quantification was validated using the NIST/ISMRM Quantitative MRI phantom, and T1 and T2 values of white matter, gray matter, and cerebrospinal fluid were measured in 19 healthy subjects. To assess MRF performance in the lower SNR regime of 0.55 T, measurement precision was calculated from 100 simula

SUBMITTER: Campbell-Washburn AE 

PROVIDER: S-EPMC8749356 | biostudies-literature | 2021 Sep

REPOSITORIES: biostudies-literature