Unknown

Dataset Information

Low-cost electrochemical detection of arsenic in the groundwater of Guanajuato state, central Mexico using an open-source potentiostat.


ABSTRACT: Arsenic is a carcinogenic groundwater contaminant that is toxic even at the parts-per-billion (ppb) level and its on-site determination remains challenging. Colorimetric test strips, though cheap and widely used, often fail to give reliable quantitative data. On the other hand, electrochemical detection is sensitive and accurate but considerably more expensive at the onset. Here, we present a study on arsenic detection in groundwater using a low-cost, open-source potentiostat based on Arduino technology. We tested different types of gold electrodes (screen-printed and microwire) with anodic stripping voltammetry (ASV), achieving low detection limits (0.7 μg L-1). In a study of arsenic contaminated groundwaters in Mexico, the microwire technique provides greater accuracy than test strips (r

SUBMITTER: Bullen JC 

PROVIDER: S-EPMC8769315 | biostudies-literature | 2022

REPOSITORIES: biostudies-literature