ABSTRACT: Background
Measurement of oxygen saturation (SpO2) during the 6 minute walk test (6MWT) could be impacted by the measurement site.Aims
To compare SpO2 and heart rate (HR) between forehead and finger sensors during the 6MWT. Sensor readings were also to be compared for signal quality and with capillary blood gas (CBG) pre and post 6MWT.Method
80 subjects with pulmonary vascular disease (PVD) and/or interstitial lung disease (ILD) performed the 6MWT. Pulse oximetry was recorded at 30 s intervals. CBG was taken pre and post 6MWT to determine capillary oxygen saturation (SCO2).Results
The forehead sensor recorded higher values for SpO2 (p < 0.001) and HR (p < 0.