Virtual bronchoscopy-guided lung SAbR: dosimetric implications of using AAA versus Acuros XB to calculate dose in airways.
Ontology highlight
ABSTRACT: In previous works, we showed that incorporating individual airways as organs-at-risk (OARs) in the treatment of lung stereotactic ablative radiotherapy (SAbR) patients potentially mitigates post-SAbR radiation injury. However, the performance of common clinical dose calculation algorithms in airways has not been thoroughly studied. Airways are of particular concern because their small size and the density differences they create have the potential to hinder dose calculation accuracy. To address this gap in knowledge, here we investigate dosimetric accuracy in airways of two commonly used dose calculation algorithms, the anisotropic analytical algorithm (AAA) and Acuros-XB (AXB), recreating clinical treatment plans on a cohort of four SAbR patients. A virtual bronchoscopy software was used