Aortic valve imaging using 18F-sodium fluoride: impact of triple motion correction.


ABSTRACT:

Background

Current 18F-NaF assessments of aortic valve microcalcification using 18F-NaF PET/CT are based on evaluations of end-diastolic or cardiac motion-corrected (ECG-MC) images, which are affected by both patient and respiratory motion. We aimed to test the impact of employing a triple motion correction technique (3 × MC), including cardiorespiratory and gross patient motion, on quantitative and qualitative measurements.

Materials and methods

Fourteen patients with aortic stenosis underwent two repeat 30-min PET aortic valve scans within (29 ± 24) days. We considered three different image reconstruction protocols; an end-diastolic reconstruction protocol (standard) utilizing 25% of the acquired data, an ECG-gated (four ECG gates) reconstruction (ECG