A system for real-time monitoring of breath-hold via assessment of internal anatomy in tangential breast radiotherapy.
Ontology highlight
ABSTRACT: The deep inspiration breath-hold (DIBH) technique assists in sparing the heart, lungs, and liver during breast radiotherapy (RT). The quality of DIBH is currently assessed via surrogates which correlate to varying degrees with the patient's internal anatomy. Since modern linacs are equipped with an electronic portal imaging device (EPID), images of the irradiated anatomy streamed from EPIDs and analyzed in real time could significantly improve assessment of the quality of DIBH. A system has been developed to quantify the quality of DIBH during tangential breast RT by analyzing the "beam's eye view" images of the treatment fields. The system measures the lung depth (LD) and the distance from the breast surface to the posterior tangential radiation field edge (skin distance, SD) at three use