Thrombin generation measured on ST Genesia, a new platform in the coagulation routine lab: Assessment of analytical and between-subject variation.


ABSTRACT:

Background

The thrombin generation (TG) assay, which measures global coagulation, has mainly been used as a research tool to investigate thrombotic and bleeding disorders. Recently, Diagnostica Stago launched the ST Genesia, a fully automated system to perform "routine version" of this assay. The objectives of this study were to evaluate the imprecision compared with the previous method, Thrombinoscope CAT, and to establish reference intervals.

Methods

Thrombin generation was measured in platelet-poor citrated plasma from 20 normal controls (fresh and after freezing at -80°C up to 12-13 weeks) on CAT and ST Genesia in duplicate to estimate the total variation, and within and between variations. The reference intervals were estimated nonparametrically in 30 men, 30 women taki

SUBMITTER: Kristensen SR 

PROVIDER: S-EPMC8803999 | biostudies-literature | 2022 Jan