Dynamic oxygenator blood volume during prolonged extracorporeal life support.


ABSTRACT: Current methods for identification of oxygenator clotting during prolonged extracorporeal life support include visual inspection, evaluation of oxygenator resistance and oxygen exchange performance, and assessment of clotting-related laboratory parameters. However, these observations do not provide a quantitative assessment of oxygenator clot formation. By measuring changes in the dynamic oxygenator blood volume this study aimed to evaluate the relation to oxygenator resistance and oxygen transfer performance. Sixty-seven oxygenators were studied during adult extracorporeal life support. Oxygenator blood volume, oxygenator resistance, and oxygen transfer efficiency were monitored. Oxygenator blood volume decreased with increasing runtime (r = -0.462; p <0.001). There was a statistically si