ABSTRACT: Purpose
The sensitivity of pseudo-continuous arterial spin labeling (PCASL) to off-resonance effects (ΔB0 ) is a major limitation at ultra-high field (≥7T). The aim of this study was to assess the effectiveness of different PCASL ΔB0 compensation methods at 7T and measure the labeling efficiency with off-resonance correction.Theory and methods
Phase offset errors induced by ΔB0 at the feeding arteries can be compensated by adding an extra radiofrequency (RF) phase increment and transverse gradient blips into the PCASL RF pulse train. The effectiveness of an average field correction (AVGcor), a vessel-specific field-map-based correction (FMcor) and a vessel-specific prescan-based correction (PScor) were compared at 7T. After correction, the P