Unknown

Dataset Information

Vibration as a pitfall in pyrosequencing analyses.


ABSTRACT: Since methylation analysis has become an important tool in forensic genetics, the reliability and credibility of the method must be ensured. After a successful validation and establishment of several pyrosequencing assays using a PyroMark® Q48 Autoprep instrument (Qiagen, Hilden, Germany), we decided to expand the method further purchasing a second instrument. But after initializing this second instrument side by side with the first, the majority of analyses failed (97 samples of 133 samples (73%)). The number of error messages increased rapidly and the average RFU values decreased. After purchasing two anti-vibration weighing tables for the PyroMark® instruments and repeating the analyses under the same conditions and with identical samples the results improved considerably, 115 samples o

SUBMITTER: Konrad H 

PROVIDER: S-EPMC8813862 | biostudies-literature | 2022 Jan

REPOSITORIES: biostudies-literature