Direct detection of Legionella species from bronchoalveolar lavage and open lung biopsy specimens: comparison of LightCycler PCR, in situ hybridization, direct fluorescence antigen detection, and culture.


ABSTRACT: We developed a rapid thermocycling, real-time detection (also known as real-time PCR) method for the detection of Legionella species directly from clinical specimens. This method uses the LightCycler (Roche Molecular Biochemicals, Indianapolis, Ind.) and requires approximately 1 to 2 h to perform. Both a Legionella genus PCR assay and Legionella pneumophila species-specific PCR assay were designed. A total of 43 archived specimens from 35 patients were evaluated, including 19 bronchoalveolar lavage (BAL) specimens and 24 formalin-fixed, paraffin-embedded open lung biopsy specimens. Twenty-five of the specimens were culture-positive for Legionella (9 BAL specimens and 16 tissue specimens). BAL specimens were tested by LightCycler PCR (LC-PCR) methods and by a direct fluorescent antibody (DF