ABSTRACT: Objectives
Intravenous fluid warming devices with surface heating systems transfer heat using aluminum blocks, which if uncoated elute toxic levels of aluminum into the infusate. This study examined extractable aluminum detected from prolonged use of the updated version of the enFlow® cartridge, which uses a parylene-coated aluminum heating block.Methods
In dynamic bench tests, we measured the concentration of aluminum that leached into three solutions (Sterofundin ISO, Plasma-Lyte 148, and whole blood) that were continuously pumped (0.2 and 5.5 mL min-1) and warmed to 40°C by the enFlow cartridge (parylene-coated) for 5 h. Prolonged quasi-static bench tests measured aluminum concentration in 16 solutions which were gently rocked within the enFlow cart