Toward a Servoregulation Controller to Automate CO2 Removal in Wearable Artificial Lungs.
Ontology highlight
ABSTRACT: A laptop-driven, benchtop control system that automatically adjusts carbon dioxide (CO2) removal in artificial lungs (ALs) is described. The proportional-integral-derivative (PID) feedback controller modulates pump-driven air sweep gas flow through an AL to achieve a desired exhaust gas CO2 partial pressure (EGCO2). When EGCO2 increases, the servoregulator automatically and rapidly increases sweep flow to remove more CO2. If EGCO2 decreases, the sweep flow decreases to reduce CO2 removal. System operation was tested for 6 hours in vitro using bovine blood and in vivo in three proof-of-concept sheep experiments. In all studies, the controller automatically adjusted the sweep gas flow to rapidly (<1 minute) meet the specified EGCO2 level when challenged with changes in inlet blood or target
SUBMITTER: Potkay JA
PROVIDER: S-EPMC8828797 | biostudies-literature | 2022 May
REPOSITORIES: biostudies-literature