Technical evaluation of the cone-beam computed tomography imaging performance of a novel, mobile, gantry-based X-ray system for brachytherapy.


ABSTRACT:

Purpose

A novel, mobile cone-beam computed tomography (CBCT) system for image-guided adaptive brachytherapy was recently deployed at our hospital as worldwide first site. Prior to the device's clinical operation, a profound characterization of its imaging performance was conducted. This was essential to optimize both the imaging workflow and image quality for achieving the best possible clinical outcomes. We present the results of our investigations.

Methods

The novel CBCT-system features a ring gantry with 121 cm clearance as well as a 43.2 × 43.2 cm2 flat-panel detector, and is controlled via a tablet-personal computer (PC). For evaluating its imaging performance, the geometric reproducibility as well as imaging fidelity, computed tomography (CT)-number accuracy

SUBMITTER: Karius A 