Simultaneous measurement of inhaled air and exhaled breath by double multicapillary column ion-mobility spectrometry, a new method for breath analysis: results of a feasibility study.


ABSTRACT: The high sensitivity of the methods applied in breath analysis entails a high risk of detecting analytes that do not derive from endogenous production. Consequentially, it appears useful to have knowledge about the composition of inhaled air and to include alveolar gradients into interpretation. The current study aimed to standardise sampling procedures in breath analysis, especially with multicapillary column ion-mobility spectrometry (MCC-IMS), by applying a simultaneous registration of inhaled air and exhaled breath. A "double MCC-IMS" device, which for the first time allows simultaneous analysis of inhaled air and exhaled breath, was developed and tested in 18 healthy individuals. For this, two BreathDiscovery instruments were coupled with each other. Measurements of inhaled air and ex