Unknown

Dataset Information

Magnetic resonance angiography and perfusion mapping by arterial spin labeling using Fourier transform-based velocity-selective pulse trains: Examination on a commercial perfusion phantom.


ABSTRACT:

Purpose

Benchmarking of flow and perfusion MR techniques on standardized phantoms can facilitate the use of advanced angiography and perfusion-mapping techniques across multiple sites, field strength, and vendors. Here, MRA and perfusion mapping by arterial spin labeling (ASL) using Fourier transform (FT)-based velocity-selective saturation and inversion pulse trains were evaluated on a commercial perfusion phantom.

Methods

The FT velocity-selective saturation-based MRA and FT velocity-selective inversion-based ASL perfusion imaging were compared with time-of-flight and pseudo-continuous ASL at 3 T on the perfusion phantom at two controlled flow rates, 175 mL/min and 350 mL/min. Velocity-selective MRA (VSMRA) and velocity-selective ASL (VSASL) were each performed with three

SUBMITTER: Xu F 

PROVIDER: S-EPMC8861891 | biostudies-literature | 2021 Sep

REPOSITORIES: biostudies-literature