Generation of ozone during irradiation using medical linear accelerators: an experimental study.


ABSTRACT:

Background

Some patients have noted a foul odor during radiation therapy sessions, but the cause of the odor remains unknown. Since we suspected that this phenomenon is due to ozone generated by ionizing radiation, this experimental study measured ozone concentrations in the treatment room and in a coiled polyvinyl chloride (PVC) tube placed within the radiation field.

Methods

We measured ozone concentrations using an ultraviolet absorption method and an ozone monitor. A PVC tube (inner diameter 7 mm, outer diameter 10 mm) was used to mimic the environment of the nasal cavity. The tube (790 cm) was coiled and set between two 4-cm-thick (for X-rays) or 2-cm-thick (for electron beams) water-equivalent solid phantoms. The sampling tube of the ozone monitor was inserted into the