Unknown

Dataset Information

Comparison of a New Optical Biometer That Combines Scheimpflug Imaging With Partial Coherence Interferometry With That of an Optical Biometer Based on Swept-Source Optical Coherence Tomography and Placido-Disk Topography.


ABSTRACT:

Purpose

To evaluate measurement precision and to compare the Pentacam AXL (Oculus Optikgeräte, Wetzlar, German), a new optical biometer based on Scheimpflug imaging and partial coherence interferometry (PCI) with that of the OA-2000 biometer (Tomey, Nagoya, Japan), which combines swept-source optical coherence tomography (SS-OCT) and Placido-disk topography.

Methods

Axial length (AL), central corneal thickness (CCT), anterior chamber depth (ACD), aqueous depth (AQD), mean keratometry (Km), astigmatism vectors J0, J45, and corneal diameter (CD) were measured in triplicate by two technical operators. Within-subject standard deviation (Sw), repeatability and reproducibility (2.77 Sw), coefficient of variation (CoV), and intraclass correlation coefficient (ICC) were used to asse

SUBMITTER: Chen S 

PROVIDER: S-EPMC8866319 | biostudies-literature | 2021

REPOSITORIES: biostudies-literature