Long-Term Stability of Gradient Characteristics Warrants Model-Based Correction of Diffusion Weighting Bias.


ABSTRACT: The study aims to test the long-term stability of gradient characteristics for model-based correction of diffusion weighting (DW) bias in an apparent diffusion coefficient (ADC) for multisite imaging trials. Single spin echo (SSE) DWI of a long-tube ice-water phantom was acquired quarterly on six MR scanners over two years for individual diffusion gradient channels, along with B0 mapping, as a function of right-left (RL) and superior-inferior (SI) offsets from the isocenter. Additional double spin-echo (DSE) DWI was performed on two systems. The offset dependences of derived ADC were fit to 4th-order polynomials. Chronic shim gradients were measured from spatial derivatives of B0 maps along the tube direction. Gradient nonlinearity (GNL) was modeled using vendor-provided gradient field des

SUBMITTER: Pang Y 

PROVIDER: S-EPMC8875771 | biostudies-literature | 2022 Feb