Impact of Increased End-Tidal Carbon Dioxide on Continuous Noninvasive Hemoglobin Monitoring during Laparoscopic Gastrectomy: A Randomized Controlled Study.
Ontology highlight
ABSTRACT: The pulse CO-Oximetry allows continuous, noninvasive monitoring of hemoglobin (SpHb). We assessed the impact of increased end-tidal carbon dioxide (EtCO2) on the accuracy and trending ability of SpHb in laparoscopic surgery. Participants (n = 64) were randomly allocated to the low carbon dioxide (CO2) group (EtCO2: 30-35 mmHg) or the high CO2 group (EtCO2: 40-45 mmHg). The SpHb and laboratory hemoglobin (tHb) were obtained during surgery. The correlation coefficient (r) between SpHb and tHb showed greater tendency in the low CO2 group (r = 0.68) than in the high CO2 group (r = 0.43). The bias (precision) was -1.18 (1.09) with a limit of agreement (LOA) of -3.31 to 0.95 in low CO2 group and -1.02 (1.24) with a LOA of -3.45 to 1.42 in high CO2 group; they did not differ significantly between
SUBMITTER: Kim HY
PROVIDER: S-EPMC8876483 | biostudies-literature | 2022 Jan
REPOSITORIES: biostudies-literature