Clinical consequences of consecutive self-expanding transcatheter heart valve iterations.


ABSTRACT:

Objective

To compare early clinical outcomes after transcatheter aortic valve implantation (TAVI) with three consecutive generations of self-expanding valves (SEVs).

Methods

Clinical endpoints of consecutive patients who underwent TAVI with CoreValve, Evolut R or Evolut PRO were included in a prospective database.

Results

TAVI was performed with CoreValve (n = 116), Evolut R (n = 160) or Evolut PRO (n = 92). Evolut R and Evolut PRO showed a tendency towards lower permanent pacemaker implantation (PPI) rates compared to CoreValve (CoreValve 27% vs Evolut R 16% vs Evolut PRO 18%, p = 0.091). By multivariable regression analysis CoreValve had a significantly higher risk for PPI (odds ratio (OR) 2.79, 95% confidence interval (CI) 1.31-5.94, p = 0.008) compared to Evolut R, while Evolut R and PRO were similar. Severe paravalvular leakage (PVL) occurred only with CoreValve, but no significant difference was observed in moderate PVL (10% vs 8% vs 6%, p = 0.49). CoreValve had a tendency towards a higher risk for more-than-mild PVL as compared with the Evolut platform (R + PRO) (OR 2.46, 95% CI 0.98-6.16, p = 0.055). No significant differences in all-cause mortality (7% vs 4% vs 1%, p = 0.10), stroke (6% vs 3% vs 2%, p = 0.21) or major vascular complications (10% vs 12% vs 4%, p = 0.14) were observed.

Conclusions

TAVI with self-expanding valves was safe, and device iterations may result in a lower need for PPI. More-than-mild PVL seemed to occur less often with repositionable technology.

SUBMITTER: Kroon HG 

PROVIDER: S-EPMC8881514 | biostudies-literature | 2022 Mar

REPOSITORIES: biostudies-literature

altmetric image

Publications

Clinical consequences of consecutive self-expanding transcatheter heart valve iterations.

Kroon H G HG   van Gils L L   Ziviello F F   van Wiechen M P H MPH   Ooms J F W JFW   Rahhab Z Z   El Faquir N N   Maugenest A-M AM   Goudzwaard J A JA   Cummins P P   Lenzen M M   Kardys I I   Daemen J J   Mattace-Raso F F   de Jaegere P P T PPT   Van Mieghem N M NM  

Netherlands heart journal : monthly journal of the Netherlands Society of Cardiology and the Netherlands Heart Foundation 20210429 3


<h4>Objective</h4>To compare early clinical outcomes after transcatheter aortic valve implantation (TAVI) with three consecutive generations of self-expanding valves (SEVs).<h4>Methods</h4>Clinical endpoints of consecutive patients who underwent TAVI with CoreValve, Evolut R or Evolut PRO were included in a prospective database.<h4>Results</h4>TAVI was performed with CoreValve (n = 116), Evolut R (n = 160) or Evolut PRO (n = 92). Evolut R and Evolut PRO showed a tendency towards lower permanent  ...[more]

Similar Datasets

| S-EPMC6899934 | biostudies-literature
| S-EPMC8301021 | biostudies-literature
| S-EPMC10874896 | biostudies-literature
| S-EPMC9917710 | biostudies-literature
| S-EPMC9581280 | biostudies-literature
| S-EPMC10205599 | biostudies-literature
| S-EPMC8302783 | biostudies-literature
| S-EPMC10157103 | biostudies-literature
| S-EPMC10774812 | biostudies-literature
| S-EPMC10119480 | biostudies-literature