Unknown

Dataset Information

Feasibility of Single-Time-Point Dosimetry for Radiopharmaceutical Therapies.


ABSTRACT: Because of challenges in performing routine personalized dosimetry in radiopharmaceutical therapies, interest in single-time-point (STP) dosimetry, particularly using only a single SPECT scan, is on the rise. Meanwhile, there are questions about the reliability of STP dosimetry, with limited independent validations. In the present work, we analyzed 2 STP dosimetry methods and evaluated dose errors for several radiopharmaceuticals based on effective half-life distributions. Methods: We first challenged the common assumption that radiopharmaceutical effective half-lives across the population are gaussian-distributed (i.e., follow a normal distribution). Then, dose accuracy was estimated using 2 STP dosimetry methods for a wide range of potential post injection (p.i.) scan time points

SUBMITTER: Hou X 

PROVIDER: S-EPMC8882881 | biostudies-literature | 2021 Jul

REPOSITORIES: biostudies-literature