Robotic assistance for percutaneous needle insertion in the kidney: preclinical proof on a swine animal model.


ABSTRACT:

Background

We evaluated the accuracy, safety, and feasibility of a computed tomography (CT)-guided robotic assistance system for percutaneous needle placement in the kidney.

Methods

Fiducials surgically implanted into the kidneys of two pigs were used as targets for subsequent robotically-assisted needle insertion. Robotically-assisted needle insertions and CT acquisitions were coordinated using respiratory monitoring. An initial scan volume data set was used for needle insertion planning defining skin entry and target point. Then, needle insertion was performed according to robot positioning. The accuracy of needle placement was evaluated upon the distance between the needle tip and the predefined target on a post needle insertion scan. A delayed contrast-enhanced CT scan w

SUBMITTER: de Baere T 

PROVIDER: S-EPMC8901810 | biostudies-literature | 2022 Mar