Vulnerabilities of radiomic features to respiratory motion on four-dimensional computed tomography-based average intensity projection images: A phantom study.


ABSTRACT:

Purpose

To evaluate the influence of respiratory motion on the robustness of radiomic features on four-dimensional computed tomography (4DCT)-based average intensity projection (AIP) images by employing an anthropomorphic chest phantom.

Methods

Three spherical objects (φ30 mm), namely, acrylic (100 Hounsfield unit [HU], homogeneous), rubber (-140 HU, homogeneous), and cork (-630 HU, heterogeneous), were moved with motion amplitudes of 0, 1, 2.5, 4, 6, 8, and 10 mm in the phantom, and 4DCT scans were repeated at four different locations. Thereafter, the AIP images were generated considering the average of the 10 respiratory phases of the 4DCT images. Further, the targets were manually delineated on the AIP images in the lung window setting. A total of 851 radiomic features, i