ABSTRACT: Background
To investigate the efficacy of the 135° hip screw, 95° intramedullary hip screw (IMHS) and 95° hip screw in the treatment of intertrochanteric reverse dip fracture of the femur.Methods
We retrospectively analyzed 125 matched pairs of human femurs (median age 64 years) which were osteotomized at a 33° angle in the left femur and extended downward from the minor trochanter to simulate a reverse oblique intertrochanteric fracture. The right femur served as a control. The left femur (n=4) was implanted with a 135° hip screw, 95° hip screw, or IMHS. A strain detector was placed distal to the fracture site to monitor fragment strain. The lateral displacement of the proximal femur was measured by a linear variable differential transformer. An Instron tester measured sti