ABSTRACT: Background
Wearable continuous monitoring biosensor technologies have the potential to transform postoperative care with early detection of impending clinical deterioration.Objective
Our aim was to validate the accuracy of Cloud DX Vitaliti continuous vital signs monitor (CVSM) continuous noninvasive blood pressure (cNIBP) measurements in postsurgical patients. A secondary aim was to examine user acceptance of the Vitaliti CVSM with respect to comfort, ease of application, sustainability of positioning, and aesthetics.Methods
Included participants were ≥18 years old and recovering from surgery in a cardiac intensive care unit (ICU). We targeted a maximum recruitment of 80 participants for verification and acceptance testing. We also oversampled to minimize the effec