Dual energy CT arthrography in shoulder instability: successful iodine removal with virtual non-contrast images and accurate 3D reformats of the glenoid for assessment of bone loss.


ABSTRACT:

Objective

To evaluate the image quality of dual energy CT (DECT) of the shoulder after arthrography and of virtual non-contrast (VNC) 3D reformats of the glenoid and to compare glenoid measurements on VNC 3D reformats and on 2D CTs.

Materials and methods

DECT arthrography (80 kV/140 kV) was performed in 42 shoulders of 41 patients with instability using diluted iodinated contrast media (80 mg/ml). VNC images and VNC 3D reformats of the glenoid were calculated using image postprocessing. Dose parameters, CT values of intraarticular iodine and muscle, image contrast (iodine/muscle), and image quality (5-point scale: 1 = worst, 5 = best) were evaluated. Two independent readers assessed glenoid morphology and performed glenoid measurements on 2D and 3D images.

Results

Cal