Atrial and ventricular flows across a transcatheter mitral valve.


ABSTRACT:

Objectives

The objective of this study was to evaluate the haemodynamic performance of transcatheter mitral valve replacement (TMVR) Implant with a focus on turbulence and washout adjacent to the ventricular surface of the leaflets. TMVR holds the promise of treating a large spectrum of mitral valve diseases. However, the haemodynamic performance and flow dynamics of such replacements are not fully understood.

Methods

A tri-leaflet biopsrosthetic TMVR represented by Caisson implant of size 36A was implanted in the mitral position of a left heart simulator pulse duplicating system under physiological conditions. The 36A implant covers an anterior-posterior range of 26-32 mm and a commissure-to-commissure range of 30-36 mm. Transmitral pressure gradient, effective orifice area