Evaluation of ocular biometry in primary angle-closure disease with two swept source optical coherence tomography devices.


ABSTRACT:

Purpose

To investigate agreement between 2 swept source OCT biometers, IOL Master700 and Anterion, in various ocular biometry and intraocular lens (IOL) calculations of primary angle-closure disease (PACD).

Setting

Rajavithi Hospital, Bangkok, Thailand.

Design

Prospective comparative study.

Methods

This study conducted in a tertiary eye care center involving biometric measurements obtained with 2 devices in phakic eye with diagnosis of PACD. Mean difference and intraclass correlation coefficient (ICC) with confidence limits were assessed, and calculations of estimated residual refraction of the IOL were analysed using Barrett's formula.

Results

Sixty-nine eyes from 45 PACD patients were enrolled for the study. Excellent agreement of various parameters wa