Field Evaluation and Calibration of Low-Cost Air Pollution Sensors for Environmental Exposure Research.


ABSTRACT: This paper seeks to evaluate and calibrate data collected by low-cost particulate matter (PM) sensors in different environments and using different aggregated temporal units (i.e., 5-s, 1-min, 10-min, 30 min intervals). We first collected PM concentrations (i.e., PM1, PM2.5, and PM10) data in five different environments (i.e., indoor and outdoor of an office building, a train platform and lobby of a subway station, and a seaside location) in Hong Kong, using five AirBeam2 sensors as the low-cost sensors and a TSI DustTrak DRX Aerosol Monitor 8533 as the reference sensor. By comparing the collected PM concentrations, we found high linearity and correlation between the data reported by the AirBeam2 sensors in different environments. Furthermore, the results suggest that the accuracy and bias

SUBMITTER: Huang J 

PROVIDER: S-EPMC8948698 | biostudies-literature | 2022 Mar