Mechanical power thresholds during mechanical ventilation: An experimental study.


ABSTRACT: The extent of ventilator-induced lung injury may be related to the intensity of mechanical ventilation--expressed as mechanical power. In the present study, we investigated whether there is a safe threshold, below which lung damage is absent. Three groups of six healthy pigs (29.5 ± 2.5 kg) were ventilated prone for 48 h at mechanical power of 3, 7, or 12 J/min. Strain never exceeded 1.0. PEEP was set at 4 cmH2 O. Lung volumes were measured every 12 h; respiratory, hemodynamics, and gas exchange variables every 6. End-experiment histological findings were compared with a control group of eight pigs which did not undergo mechanical ventilation. Functional residual capacity decreased by 10.4% ± 10.6% and 8.1% ± 12.1% in the 7 J and 12 J groups (p = 0.017, p < 0.001) but not in the

SUBMITTER: Romitti F 

PROVIDER: S-EPMC8957661 | biostudies-literature | 2022 Mar

REPOSITORIES: biostudies-literature