Unknown

Dataset Information

Amperometric immunosensor developed for sensitive detection of SARS-CoV-2 spike S1 protein in combined with portable device.


ABSTRACT: In this present study, an amperometric immunosensor was developed based on disposable screen-printed carbon electrode (SPCE) for specific and sensitive detection of SARS-CoV-2 S1 protein. Anti-SARS-CoV-2 S1 monoclonal antibody was firstly immobilized onto the electrode surface. Then, the sandwich complex was formed by addition of S1 protein, secondary antibody and HRP-IgG, respectively. Chronoamperometry measurements were done in the presence of TMB mediator and the detection of SARS-CoV-2 S1 protein was performed by using 10 μL sample. The limit of detection (LOD) was found to be 0.19 ng/mL (equals to 24.7 amol in 10 μL sample) in the linear range of 0.5-10 ng/mL obtained in buffer medium. The applicability of this assay was investigated in the linear range of 0.5-3 ng/mL S1 protein in ar

SUBMITTER: Erdem A 

PROVIDER: S-EPMC8963973 | biostudies-literature | 2022 Jul

REPOSITORIES: biostudies-literature