ABSTRACT: Purpose
Daily clinical use of therapeutic light sources can lead to changes in light emission stability with potentially significant consequences for usage in photomedicine treatment. The aim of this study was to evaluate the average and maximum power and to describe the beam diameter of different low-power laser photobiomodulation devices in clinical use in Brazil.Methods
The power and light-emitting beam diameter of twenty-four therapeutic devices with an average age of 11±5 years, with an average weekly use of fewer than thirty minutes, were measured.Results
The analyzed power varied between 2% to 134% of the values declared by the manufacturers. Differences in beam diameter of between 38% and 543% of the nominal values were also observed. It is also noteworthy t