Comparative hard x-ray tomography for virtual histology of zebrafish larva, human tooth cementum, and porcine nerve.
Ontology highlight
ABSTRACT: Purpose: Synchrotron radiation-based tomography yields microanatomical features in human and animal tissues without physical slicing. Recent advances in instrumentation have made laboratory-based phase tomography feasible. We compared the performance of three cutting-edge laboratory systems benchmarked by synchrotron radiation-based tomography for three specimens. As an additional criterion, the user-friendliness of the three microtomography systems was considered. Approach: The three tomography systems-SkyScan 2214 (Bruker-microCT, Kontich, Belgium), Exciscope prototype (Stockholm, Sweden), and Xradia 620 Versa (Zeiss, Oberkochen, Germany)-were given 36 h to measure three medically relevant specimens, namely, zebrafish larva, archaeological human tooth, and porcine nerve. Th
SUBMITTER: Migga A
PROVIDER: S-EPMC8968075 | biostudies-literature | 2022 May
REPOSITORIES: biostudies-literature
ACCESS DATA