Unknown

Dataset Information

Development of an independent MU calculation software for radiotherapy treatments with stereotactic cones.


ABSTRACT:

Purpose

Development of an independent MU calculator (StereoCalc) with and without heterogeneity corrections for stereotactic treatments, in a Varian TrueBeam STx LINAC using stereotactic cones, with flattening filter-free photon energies.

Methods

Multiple depth curves and output factors were measured, following the dosimetry formalism for small fields proposed by the TRS-483. The developed StereoCalc imports and processes the beam data files and calculates the patient plans with and without heterogeneity correction. Validation of the developed software was carried out using phantoms. The accuracy of the StereoCalc software was verified in stereotactic patient plans.

Results

A maximum difference of 2.47% and 2.07% was obtained in the phantom validation tests with and w

SUBMITTER: Campos GFP 

PROVIDER: S-EPMC8992931 | biostudies-literature | 2022 Apr

REPOSITORIES: biostudies-literature