Ablation Index Outcome in Redo Persistent Atrial Fibrillation Ablation: Results of a Short-Term Study.


ABSTRACT:

Background

Ablation index (AI) is a novel catheter-based parameter that has improved the outcome and safety of radiofrequency (RF) ablation of pulmonary vein isolations (PVIs). This index incorporates contact force (CF) (g), time (s), and power (W) parameters. The role of AI in redo ablations for persistent atrial fibrillation (peAF) has not been fully investigated. Hence, the impact of AI on the success of the redo PVI during the short-term follow-up period is the aim of this study.

Methods

A retrospective analysis of 39 consecutive patients who underwent redo PVI ablations for peAF was carried out between January 2016 and December 2018. Target values for AI were 500 - 550 for anterior and roof and 400 - 380 for posterior and inferior regions. We compared outcomes between A